Your Reliable Partner for THC Measurements

Total Hydrocarbon Analyzer RS53-T

Applications

  • Emission Monitoring
  • Optimizing Processes
  • Combustion Control
  • Threshold Limit Values (TLV) - Measurements
Total Hydrocarbon Analyzer RS53-T Front

Features

  • Flame Ionization Detector (FID)
  • Continuous Measurement Mode
  • Dual Channel Analyzer available (RS53-T/2)
  • TUV-approved (meets EPA-approval)
  • Heated Analytical Compartment (up to 200 °C / 473 K)
  • Auto Ignition
  • Flame-Out Indication
  • Automatic Hydrogen Cut-Off
  • Self-Monitoring Capabilities
  • Electronic SAmple Pressure Regulator
  • Auto / Manual Range
  • Zero and Calibration over Sample Line
  • Internal Zero-Gas and Combustion-Air Generator
  • Integrated Sample Filter
  • Integrated Fuel Gas Cylinder
  • Integrated Calibration Gas Cylinder
  • Hydrogen / Helium-Hydrogen Fuel Selector
  • Easy Handling
  • Portable (approx. 20 kg)
  • Option: ppb-Measurement Ranges
  • Option: Sample Lines
  • Option: External Filters
  • Option: USB-Interface
  • Option: Filling Adaptors

Specifications

(Reference Quantity: Full Range Deflection = FRD)

5 Measurement Ranges
  • 0-10 ppm Propane
  • 0-100 ppm Propane
  • 0-1.000 ppm Propane
  • 0-10.000 ppm Propane
  • 0-100.000 ppm Propane
Linearity +/- 2 % MBE (up to Range 4)
Oxygen Cross Sensitivity < +/- 2 % FRD (with He/H2 Fuel)
Limit of Detection < 2 % FRD
Zero Driftt < 2 % FRD/week
Span Drift < 4 % FRD/week
Response Time90 approx. 1-2 s
Sample Flow Rate approx. 90 l/h
Analytical Temperature up to 200°C
Signal Output 0-10 V and/or 0/4-20 mA
Status Contacts
  • Concentration-Alarm
  • Range Identification
  • Flame Out Indication
  • Sample Flow Alarm
Dimensions 360 x 200 x 500 mm (WxHxD)
Weight approx. 20 kg
Ambient Temperature +5 ... +35 °C
Main Power 230 V / 50 Hz or 115 V / 60(50) Hz / 450 Watts